Untuned broadband spiral micro-coils achieve sensitive multi-nuclear NMR TX/RX from microfluidic samples.

The low frequency plateau in the frequency response of an untuned micro-resonator permits broadband radio-frequency reception, albeit at the expense of optimal signal-to-noise ratio for a particular nucleus. In this contribution we determine useful figures of merit for broadband micro-coils, and the...

Full description

Saved in:
Bibliographic Details
Published in:Scientific Reports Vol. 11; no. 1; pp. 1 - 13
Main Authors:
Format: Article
Published: Springer Nature, 4/8/2021
Subjects:
Online Access:Go to the source
Tags: Add Tag
Be the first to tag this record!
Be the first to leave a comment!
You must be logged in first